Witryna9 lut 2024 · The Impella percutaneous microaxial ventricular assist device provides significant ventricular unloading, but at the cost of increased major bleeding, … Witryna7 kwi 2024 · The use of sodium bicarbonate purge solution in an Impella device is described in two patients with heparin-induced thrombocytopenia (HIT). The first case details a patient with severe mitral regurgitation and cardiogenic shock who had an Impella CP placed who developed HIT.
Optimizing anticoagulation for patients receiving Impella support
Witryna22 lip 2024 · The direct thrombin inhibitor (DTI) bivalirudin is often used in the setting of heparin-induced thrombocytopenia (HIT). However, Hohlfelder et al. demonstrated that when bivalirudin was used in the Impella purge solution, 7 of the 9 patients in the study experienced decreases in purge flow and increases in purge pressure, leading to the ... WitrynaQuick Reference and Troubleshooting Guide Impella® Heart Pump with Automated Impella® Controller Function Impella Heart Pump Positioning Purge System ... (a severe form of deep venous thrombosis), Pulmonary valve insufficiency, Respiratory dysfunction, Sepsis, Thrombocytopenia, Thrombotic vascular (non-central nervous system) … flitch and burl
Evaluation of Thrombocytopenia in Patients Receiving Percuta ...
WitrynaThe Impella 5.0 (Figure 1) is a microaxial, catheter-mounted pump that works on the principle of Archimedes’ screw. 8 It is able to generate a total output of 4.0-5.0 L/min at an RPM of 33,000, a step up from its predecessor, the Impella 2.5, which generates a maximum of 2.0-2.5 L/min. Witryna14 sie 2024 · Thrombocytopenia can occur when using an Impella percutaneous ventricular assist device (pVAD), and heparin-induced thrombocytopenia (HIT) … Witryna• The Impella percutaneous microaxial ventricular assist device provides significant ventricular unloading, but at the cost of increased major bleeding, hemolysis, and … flitch and burl whitehorse